PARAMETER SETTINGS

Parameter
Settings range
Language
DEU, ENG, ELL, SPA, FRA, ITA,
NLD, TUR, PLK, POR, CZE
P-Unit
hPa, mbar, cmH
Press.
4 – 20 (25
1
IPAP
EPAP – 20 (25
2
EPAP
4 hPa – IPAP
2
I-Slope
1 – 3
2
E-Slope
1 – 3
2
Frequency
5 – 30 bpm
3
I:E
20 – 80 %
3
Delay Time
0 – 20 s
4
Backup freq
ON, OFF
5
P-Min
4 – 20 hPa
6
P-Max
4 – 20 hPa
6
P-Start
4 – 20 hPa
6
I-FLEX
0 – 3 hPa/ (l/s)
E-FLEX
0 – 3 hPa/ (l/s)
Ramp
0 – 45 min
P-Ramp
4 hPa – 20 (25
Mask Test
0 – 90 s
Auto
OFF, Start/Stop, Start
Display Vt
ON, OFF
Bact.Filter
Yes, No
Brightness
0 – 100 %
only point 2 CPAP and point 2 AutoCPAP and point 2 BILEVEL ST20/25 in
1
modes CPAP
only point 2 BILEVEL ST20/25
2
only point 2 BILEVEL ST20/25 in modes BILEVEL T and ST
3
only point 2 BILEVEL ST20/25 in BILEVEL ST mode
4
only point 2 BILEVEL ST20/25 in BILEVEL S mode
5
only point 2 AutoCPAP
6
only point 2 BILEVEL ST25
7
56
Device functions
O
2
) hPa
7
) hPa
7
) hPa
7
Dependency
-
-
-
EPAP, 20 (25
) hPa
7
4 hPa,
IPAP
IPAP, EPAP,
Frequency, I:E
IPAP, EPAP,
Frequency, I:E
-
-
-
-
P-Start
P-Start
P-Min,
P-Max
-
-
-
CPAP:
Press.
APAP
:
P-Start
6
BILEVEL S, T, ST
:
2
EPAP
-
-
-
-
-
Setting steps
-
-
0.5 hPa
0.5 hPa
0.5 hPa
1 Level
1 Level
1 bpm
5%
1 s
-
0.5 hPa
0.5 hPa
0.5 hPa
1 hPa/ (l/s)
1 hPa/ (l/s)
5 min
0.5 hPa
5 s
-
-
-
10 %