EPAP

Q   MMMMMMMMMMA ;$
(only point 2 BILEVEL ST20/25 in modes BILEVEL S, T and ST)
4%/ !*1
Selection of expiration pressure (EPAP) by the physician
1. First set IPAP.
2. Use the selection keys
setting by pressing the enter key
The parameter I-Slope is set next.

I-SLOPE

Q (+,! MMMMMMMMMMMMMM=
(only point 2 BILEVEL ST20/25 in modes BILEVEL S, T and ST)
4%/ !*1
The I-Slope determines the time of pressure increase from EPAP to IPAP.
Level 1 = 10 hPa/s
Level 2 = 20 hPa/s
Level 3 = 40 hPa/s
Selection of I-Slope by the physician
1. First set IPAP and EPAP
2. Use the selection keys
setting by pressing the enter key
The parameter E-Slope is set next.
Calculation of the I-Slope time:
Settings:
IPAP = 10 hPa
EPAP = 6 hPa
IPAP – EPAP
I-Slope time =
10 hPa – 6 hPa
I-Slope time =
44
Device functions
and
and
I-Slope = 2
Frequency = 12 bpm
E-Slope = 2
I-Slope
= 0.2 s
20 hPa/s