ALARM FUNCTIONS OF THE DEVICE

MASK ALARM

If the mask slips off your face, the tube is pulled out or any other leak occurs
that cannot be compensated for, the device does not switch off automatically
but emits an acoustic and visual alarm.
The mask alarm is only active in the „Auto OFF" and „Auto Start" automatic modes.
Testing the mask alarm
Start the device in „Auto OFF" mode. The mask should be open, not on the
face. After a short time an acoustic signal sounds.

PRESSURE ALARM

Should an excessively high pressure occur during the treatment due to a hard-
ware error or other circumstance, the device emits an alarm sound and switches
the blower off.
The alarm is turned off by pressing the ON/OFF key.

WAKE UP ALARM

Ex factory, the alarm clock function is not enabled. But it can be unlocked
by the physician or dealer with TRENDset PC software.
If enabled, you can activate or deactivate the alarm clock with the selection
keys
and
. Pressing the ON/OFF key once mutes the alarm for the next
5 minutes and pressing it twice stops it completely.
Alarm functions of the device
57