Operation; Serial Number; Software Version; Voltage Measurement - AEMC 6505 User Manual

Megohmmeter
Hide thumbs Also See for 6505:
Table of Contents

Advertisement

Charge the instrument fully before use (see § 5.1)
4.1

Serial Number

To view the serial number of the instrument, keep the DISPLAY button
pressed and turn the switch to the MΩ-500V position.
4.2

Software Version

To view the internal software version of the instrument, keep the DISPLAY
button pressed and turn the switch to the MΩ-1000V position.
Technical documentation on Understanding Insulation Resistance Testing
is available on our website at www.aemc.com/techinfo.
4.3

Voltage Measurement

As soon as the switch is set to an insulation measurement position, the
instrument automatically measures the presence of any AC/DC voltage.
This voltage is measured at all times and indicated on the small display
unit.
The instrument automatically determines AC or DC. The AC measurement
is an RMS value.
If an excessively high external voltage is present on the terminals (> 0.4
Un), pressing the START/STOP button has no effect and no measure-
ments are made. Similarly, if an excessively high erroneous voltage (> 0.4
Un) is detected during the measurement, the measurement is automati-
cally stopped.