Command Descriptions
[SENSe]:MCPower:RCHannels:MAIN<x> (No Query Form)
Conditions
Group
Syntax
Related Commands
Arguments
Examples
[SENSe]:MCPower:RCHannels:TOTal (No Query Form)
Conditions
Group
Syntax
Arguments
Examples
2-992
Sets the power reference to the main channel with the index (<x>) in the MCPR
measurement.
The parameter <x> = 1 to 99, representing the main channel 1 to 99, respectively.
The main channel must be defined using the [SENSe]:MCPower:CHANnel:MAIN
commands.
Measurement views: MCPR
Sense commands
[SENSe]:MCPower:RCHannels:MAIN<x>
[:SENSe]:MCPower:CHANnel:MAIN commands
None
SENSE:MCPOWER:RCHANNELS:MAIN3
channel.
Sets the power reference to the total power of all the active channels in the MCPR
measurement.
Measurement views: MCPR
Sense commands
[SENSe]:MCPower:RCHannels:TOTal
None
SENSE:MCPOWER:RCHANNELS:TOTAL
of all the active channels.
RSA6100A Series & RSA5100A Series Programmer Manual
selects Main 3 for the power reference
sets the power reference to the total power
Need help?