Protocol Settings - Interacoustics AT235 Additional Information

Impedance audiometer
Hide thumbs Also See for AT235:
Table of Contents

Advertisement

AT235 Additional Information

3.2 Protocol Settings

Press the Setup button to enter protocol settings.
Note that the protocol settings of a test can only be changed while there is no data stored for that test. In
instances where data is present at the top of the screen, it will say "Data is present, press 'New session' to
allow changing settings.
The following buttons are available while in protocol settings.
Note that all changes to the settings are stored immediately. Pressing Exit will
therefore result in the settings being temporarily in use. After saving or clearing a
session by the New session button, the original protocol settings are restored.
Press Save or SaveAs to permanently save protocol settings. Factory protocols
can be temporarily saved (by pressing Exitt) or saved with a new protocol name
by pressing SaveAs. To save customized protocols under a new name, hold the
Shift button and then press SaveAs.
Moves one test up in the current protocol
Pressing shift in addition will move the selected test upwards in the protocol.
Moves one test down in the current protocol
Pressing shift in addition will move the selected test downwards in the protocol.
Use the wheel while holding the Add button to select which test is added at the
bottom of the current protocol.
Note that each added test comes with the default settings as defined in the
Instrument protocol settings.
Pressing Remove deletes the current selected test from the protocol, without
warning.
Use the wheel to browse through the different settings in the menu and press
Toggle to change check boxes, or hold Change and rotate the wheel for
changing settings with more than two options. Touching the Enter button will also
toggle between check boxes or browse through setting options.