9R479-IR
User Calibration—For qualified personnel only
meriGauge plus System User Manual
info@Meriam-Direct.com
What does "Calibration Enabled" refer to?
•
With a check in the box, Calibration Enabled displays
the Live Reading with your current user calibration
applied.
•
Without a check in the box, it displays Live Reading
without your user calibration.
Restore Factory Defaults
This button clears your user calibration from the meriSense.
Choose Calibration Unit
•
Select the calibration reference units.
•
This does not affect or depend on the configuration units.
Begin button...Apply button
1. Click
.
Begin
2. Select the point.
3. Apply reference pressure within the range shown.
4. Enter the reference value.
5. Save the point.
6. Repeat steps 2-5 until you are done.
(You have to change at least one point.)
7. Click the
button.
Apply
Note: If you click the
Back
the five boxes, an alert appears.
April 2016