User Calibration—For qualified personnel only
meriGauge plus System User Manual
info@Meriam-Direct.com
What does "Calibration Enabled" refer to?
With a check in the box, Calibration Enabled displays
the Live Reading with your current user calibration
applied.
Without a check in the box, it displays Live Reading
without your user calibration.
Restore Factory Defaults
This button clears your user calibration from the meriSense.
Choose Calibration Unit
Select the calibration reference units.
This does not affect or depend on the configuration units.
Begin button...Apply button
1. Click
.
Begin
2. Select the point.
3. Apply reference pressure within the range shown.
4. Enter the reference value.
5. Save the point.
6. Repeat steps 2-5 until you are done.
(You have to change at least one point.)
7. Click the
button.
Apply
Note: If you click the
Back
the five boxes, an alert appears.