Zero the meriGauge plus in Display mode
meriGauge plus System User Manual
info@Meriam-Direct.com
In normal measure mode
If the sensor is within a tolerance band around zero, press and
hold the
key to zero the pressure measurement and to
Zero
reset the Min and Max measurements.
Note: The tolerance band is approximately ± 1 % of the
Full Scale pressure value of the sensor.
In Min or Max mode
Press and hold the
key to reset the Min and Max
Zero
measurement. However, this does not zero the pressure
measurement.
In Tare mode
When the Tare is off (T.OFF), press and hold the
turn on Tare (T.ON) and to set the Tare value at the current
pressure measurement.
Likewise, when the Tare is on (T.ON), press and hold the
key to turn off the Tare mode.
In Average mode
Press and hold the
key to restart the rolling average.
Zero
In Data Log mode
Press and hold the
key until you see –STARTING—appear
Zero
on the LCD display to start recording a new data log.
Holding the Zero key
The key must be held to perform the Zero or Tare function.
The displayed value(s) dashes out during the zero or tare
process.
April 2016
key to
Zero
Zero