Basic setup Standard

After switching on
Switching between two point control and continuous pumping
MODE
Starting and stopping the process control
START
STOP
Venting
VENT
Attention: If pressurized
inert gas is used install an
overpressure relief valve.
Max. pressure at the
pressure transducer 1600
mbar! Ensure that high
pressure is compatible
with the mechanical
stability of the system!
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 99 90 20 / 01/12/2004
page 12 of 36
The process control is inactive, i. e. controller is ready for vacuum
control, but control operation has not been activated.
☞ The controller is in operation mode two-point control, the
actual pressure is displayed.
☞ Values for set point p and hysteresis as from last operation
are reactivated.
☞ Press key START/STOP to start process control.
☞ If using a cooling water valve, the valve must be configu-
rated, see section "How to change the basic setup".
➨ Press key MODE.
☞ Mode change is only possible if the process control is inac-
tive (after switching on or pressing key START/STOP).
☞ The two-point control display disappears or appears.
In both modes (two-point control and continuous vacuum):
➨ To start the process control: Press key START/STOP.
☞ If the cooling water valve is preselected: Cooling water valve
opens immediately, the symbol is displayed.
☞ Isolation valve opens. If a cooling water valve is preselected,
at the first demand of vacuum the isolation valve opens 2
sec. past the cooling water valve: Pumped vapour reaches
cold surfaces of condenser.
➨ To stop the process control: Press key START/STOP.
☞ Attention: Take suitable precautions if the system risks to
become pressurized due to low boiling point solvents when
the isolation valve has been closed.
☞ If a cooling water valve is preselected: Cooling water con-
tinuous to run for additional 5 min. to allow condensation,
symbol cooling water is flashing.
Short venting (e. g. in case of retardation of boiling or if sudden
foaming occurs):
➨ Press key VENT shortly (< 2 sec.).
☞ Pumping is interrupted, the process control is stopped.
Venting the system:
➨ Press key VENT continuously (> 2 sec.).
☞ Pumping is interrupted, the process control is stopped.
☞ Ventilation until pressure does not change any more, i. e.
until atmospheric pressure is attained
☞ Key STOP interrupts venting at any time.