Error messages
(continued)
Display
Source of error
code
Cover
(4 in Fig.
Safety switch triggered.
Flash standby state not reached within the pre-
scribed time. Charging electronics defective.
No flash or flash duration too short.
Flash bulb dirty or defective.
The temperature in the drying module is not
within the permissible temperature range.
Unload station mechanically blocked.
Heater fan not running or running too slowly.
No cassette is in the drying module for flashing.
Cassette left the cassette carrier, but did not
reach the drying module.
Cassette not successfully ejected from drying
module.
Cassette in the drying module during initialization
or prior to a print job being carried out.
The flap of the drying module either does not
close or does not open. The flap is blocked (e.g.
by a cassette).
Lifting table end sensor does not switch.
Lifting table position sensor does not switch dur-
ing positioning.
Flash power supply: Overtemperature for more
than 10 minutes.
Leica IP C
2) not closed properly.
7.
Troubleshooting solution
Ensure that the cover is not blocked, e.g. by the
manual unload station. Close cover completely.
Call Leica Technical Service.
Check whether flash bulb works – to do so,
observe the stray light on the cover.
Never open the cover to check whether a
flash is triggered!
No flash --> install a new flash bulb. See
Chapter
4.6.
Carry out a RESET. See
Remove the cause of the blockage.
Call Leica Technical Service.
Remove cassette.
Remove cassette.
Remove the cause of the blockage.
Remove the cause of the blockage.
Call Leica Technical Service.
Call Leica Technical Service.
Call Leica Technical Service.
Troubleshooting
Chapter
7.1.
63