Leica IP C Instructions For Use Manual page 58

Automated printing system for tissue cassettes
Hide thumbs Also See for IP C:
Table of Contents

Advertisement

6.
Cleaning and Maintenance
Instructions for storage
89
Important!
Turn the printer off and disconnect it from the power supply at this point to prevent damage to
the print head.
Never use a cleaning cartridge together with a location plate (70)!
Do not reinstall a used transport plate (73), as it will no longer completely seal the print head.
58
(continued)
70
• Then pull out the location plate (70,
• Clean the print head with a foam swab, as
described in Fig. 59 (see also
Always use the cleaning swab for the
print head once only. Never rotate the
swab—this can damage the nozzle
plate of the print head.
Fig. 64
73
Fig. 65
Instructions for Use V 1.9 RevH – 03/2017
Fig.
65).
• After cleaning the print head,
insert a new transport plate
(73).
• Push the small lever (69,
Fig.
58) back down into its
original position.
• Reclose the lid of the printer
and press any key; the print
head moves back into the rest
(park)
position.
• Screw the red cap (85, in
Fig.
57) onto the cleaning
cartridge and close the plate
on the left side again.
Fig.
64).