Eclipse 705 Installation And Operating Manual page 38

Guided wave radar level transmitter
Table of Contents

Advertisement

PROBE SPECIFICATIONS
Description
Materials
Probe
Process seal
Spacers
Probe diameter
Mounting
Process Connection
Probe length
Transition Zone ¿
Top
Bottom
Process Temp. ¡
Max
Min
Max. Process Pressure ¡
Max. Viscosity
Dielectric Range
Vacuum service
Media coating
Description
Materials
Probe
Process seal
Bottom spacer
Probe diameter
Cage
Process Connection
Measuring range
Process Temp. ¡
Max
Min
Max. Process Pressure ¡
Max. Viscosity
Dielectric Range (level only)
Vacuum service
¿
Transition Zone (zone with reduced accuracy) is dielectric dependent;
r = dielectric permitivity. It is recommended to set 4-20 mA signal out-
ε
side transition zones.
See graphs at page 39 and 40.
¡
38
7MB: standard twin rod GWR probe
316/316L (1.4401/1.4404), Hastelloy
Teflon
with Viton
GFLT, Aegis PF 128 or Kalrez
®
®
Teflon
®
Two 13 mm (0.5") Ø rods – 22,2 mm (0.875") C L to C L
In-tank mounting only.
Twin rod probe must be used in metallic vessel or stillwell > 25 mm (1") from any surface or
obstruction.
Threaded: 2" NPT or 2" BSP (G 2")
Flanged: various ANSI, EN (DIN) or torque tube mating flanges
From 60 cm to 610 cm (24" to 240")
r ≥ 1,9 = 150 mm (6")
ε
r: 1,9 = 150 mm (6")/
r: 80 = 25 mm (1")
ε
ε
+150 °C @ 27,6 bar (+300 °F @ 400 psi)
-40 °C @ 51,7 bar (-40 °F @ 750 psi)
70 bar @ +20 °C (1000 psi @ +70 °F)
1500 mPa.s (cP)
1,9 to 100
Negative pressure but not hermetic seal
Film: 3% error of coated length, bridging not recommended Æ
7EK: Top/Bottom GWR probe
r ≥ 1,4 - max +260 °C
ε
316/316L (1.4401/1.4404)
PEEK and TFE with Aegis PF 128
TFE
Inside tube: max 22,5 mm (0.88")
2" - Sch 80 Top/Bottom cage
Threaded: 1 1/2" NPT or 2" NPT
Welded: 1 1/2" or 2" socket weld
Flanged: Various ANSI, EN (DIN) or torque tube mating flanges
min 356 mm (14") Std. – max 6,1 m (240")
+260 °C @ 115 bar (+500 °F @ 1670 psi)
-15 °C @ 117 bar (0 °F @ 1700 psi)
117 bar @ -15 °C (1700 psi @ +0 °F)
10.000 mPa.s (cP)
1,4 to 100 - Non conductive and conductive
media
Negative pressure but not hermetic seal
Æ
Bridging is defined as continuous accumulation of material between the
probe elements.
C (2.4819) or Monel
(2.4360)
®
®
4079 (consult factory for alternatives)
®
7EK: Top/Bottom GWR probe
r ≥ 10 - max +315 °C
ε
PEEK and Alumina with Aegis PF 128
PEEK
+315 °C @ 109 bar (+600 °F @ 1585 psi)
10 to 100 - Conductive media