MAINTENANCE

WARNING MESSAGES

A warning message identifies the status of the instrument. A warning message does not necessarily require immediate action but may
require close attention and/or follow up. The warning message will be displayed on the unit and/or selected on the Pactware screen while
the output signal will remain normal.
Display
Action
Message
OK
None
Initial
None
DryProbe
None
WeakSgnl
1) Set transmitter to lower dielectric range
2) Increase sensitivity
Flooded?
1) Decrease level in vessel
2) Set transmitter to lower dielectric range
3) Replace with Model 7MR Overfill probe
Fid Sprd*
1) Check connection between probe and transmitter
2) Check for moisture on top of probe
3) Consult factory
Hi Temp
1) Transmitter may need to be moved to ensure ambient temperature
is within specification
2) Change to remote mount transmitter
3) Use sunshade
Lo Temp
1) Transmitter may need to be moved to ensure ambient temperature
is within specification
2) Change to remote mount transmitter
Sys Warn
Consult factory
TrimReqd
Consult factory
Cal Reqd
Consult factory
26
Comment
Normal operating mode
Program is Initializing, level
reading held at 4 mA set point.
This is a transient condition.
Normal message for a dry
probe. End of probe signal is
being detected.
Signal amplitude is lower than
desired.
Loss of level signal possibly
due to flooding, twin rod
probes only
Fiducial Ticks variation is
excessive.
Present temperature in
electronics compartment is
above +80° C
Present temperature in
electronics compartment is
below -40° C.
Unexpected but non-fatal
software event
Factory set Loop values are
defaults, loop output may be
inaccurate
Factory set default calibration
parameters are in use, level
reading may be inaccurate