Dc Voltage Specifications, Output; Dc Voltage Specifications, Isolated Input; Dc Current Specifications, Output - Omega CL3001 User Manual

Lab calibrator
Table of Contents

Advertisement

12.2 DC Voltage Specifications, Output

Absolute Uncertainty,
± (ppm of output +μV)
Ranges
1
0 to 100.000 mV
0 to 1.00000 V
0 to 10.0000 V
0 to 100.000 V
TC Output and Input
-10 to 75.000 mV 25
1. All outputs are positive only.
2. Remote sensing is not provided. Output resistance is < 1Ω
Ranges
0 to 100.000 mV
0 to 1.00000 V
0 to 10.0000 V
0 to 100.000 V

12.3 DC Voltage Specifications, Isolated Input

Ranges
0 to 10.0000 V
0 to 100.000 V

12.4 DC Current Specifications, Output

tcal ±5 °C ± (ppm of output +μA)
Ranges
1
0 to 100.000 mA
1. All outputs are positive only.
tcal ±5 °C
90 days
1 year
25
3
30
3
25
10
30
10
25
100
30
100
25
1 mV 30
1 mV
3 µV
30
3 µV
Bandwidth 0.1 to 10 Hz p-p
± (ppm of output +μV)
10 ppm+1 mV
Absolute Uncertainty,
tcal ±5 °C, ± (ppm of reading + mV)
50
50
Absolute Uncertainty,
90 days
40
1
Stability
24 hours, ±1 °C
± (ppm of output +μV)
5 ppm +2
4 ppm + 10
4 ppm + 100
5 ppm + 1 mV
5 ppm + 2 µV
Noise
1 μV
10 μV
100 μV
0.2
2.0
1 year
Resolution
50
1
1 µA
Maximum
Resolution
1 µV
10 µV
100 µV
1 mV
1 µV
Bandwidth
10 Hz to 10 kHz rms μV
6 μV
60 μV
600 μV
20 mV
Resolution
100 µV
1 mV
Maximum
Maximum
Compliance Inductive
Voltage
12 V
Burden
2
10 mA
10 mA
10 mA
1 mA
10 Ω
Load
100 mH
89

Advertisement

Table of Contents
loading