4.7 Thermocouple (T/C) Measure

The CL3001 can measure all common thermocouple types in °F or
°C, plus basic millivolts from -10.0 to 75.0 mV.
The following common thermocouple types are supported:
B, C, E, J, K, L, N, R, S, T, U, XK, BP
a) Disconnect any test leads from external devices.
b) Press the
not already selected. If RTD/O mode is displayed, press the key
again to cycle to thermocouple mode.
c) If output mode is displayed, select input mode by pressing the
and
d) Press the
the millivolt range.
e) Connect the unit under test to the thermocouple terminals of the
CL3001 using a standard T/C miniplug as shown in figure 14.
One pin is wider than the other; do not attempt to force the plug
in the wrong polarization. The T/C wire used for the connection
must match the thermocouple type selected for proper cold
junction compensation. If the CL3001 and the T/C miniplug are
key.