10.4
Remote Command Listing
The following is an alphabetical list of all CL3001 remote commands
and queries, including the common commands and the device-
dependent commands. Each command title includes a checkbox
that indicates the remote interface applicability, IEEE-488 and/or RS-
232, and the command group, Sequential or Overlapped; see
section 9.6.1 for a description of these terms.
*CLS
Sequential
Clear Status command. This command clears the ESR, the error
queue, and the RQS bit in the status byte. This command terminates
any pending operation complete commands, *OPC or *OPC?.
Parameter:
Response:
Example:
*ESE
Overlapped
Event Status Enable command. This command loads a byte into
the Event Status Enable (ESE) register. See the Event Status Enable
Register (ESE) description in section 9.7.
Parameter:
Response:
Example:
*ESE?
Overlapped
Event Status Enable query. This command returns the contents of
the Event Status Enable (ESE) register. See the Event Status Enable
Register (ESE) description in section 9.7.
Parameter:
Response:
62
IEEE-488
X
Overlapped
<None>
<None>
*CLS
This example clears the ESR, the error queue, and
the RQS bit in the status byte.
IEEE-488
X
<value>
where <value> is the decimal equivalent of the
ESE byte, 0 to 255
<None>
*ESE 140
This example loads decimal 140 (binary
10001100) to enable bits 7 (PON), 3 (DDE) and 2
(QYE).
IEEE-488
X
<None>
<value>
where <value> is the decimal equivalent of the
RS-232
X
X
RS-232
X
X
RS-232
X
X
Sequential
Sequential