7.5

RTD Transmitter

1. Disconnect any test leads from external devices.
2. Select RTD output on the primary display as described in section
4.4. Select the RTD type which corresponds to the transmitter
being tested.
3. Select current input on the isolated display as described in
section 5.2. Select the isolated loop power option. If a HART
communicator is to be used for set up of the transmitter, select
the HART option.
4. Connect the transmitter as shown in figure 24.
5. Test and calibrate the transmitter per the manufacturer's
instructions.
7.6

Thermocouple Test

Figure 25 - Thermocouple Test Application
36
Figure 24 - RTD Transmitter Application