Error Code Listing - Omega CL3001 User Manual

Lab calibrator
Table of Contents

Advertisement

Status Commands
Command
FAULT?
10.3

Error Code Listing

Error Number
1
100
101
102
103
104
105
106
108
109
110
111
112
113
114
60
Description
Returns the most recent error code in the CL3001 error
queue, and then removes that error code from the
queue.
Message Class
DDE
EXE
CME
EXE
CME
EXE
EXE
EXE
CME
CME
CME
EXE
CME
CME
CME
Description
Error queue overflow.
OPER or STBY was received when the
CL3001 is in measure mode or thermocouple
source mode.
A non-numeric entry was received in a field
that should contain a numeric entry.
The numeric field exceeds 10 characters.
Invalid units name or prefix.
An attempt to enter RTD source mode was
made when SPRT is selected, or to select
SPRT when RTD source mode is selected.
Entry is above upper limit for the selected out-
put range.
Entry is below lower limit for the selected out-
put range.
A required command parameter was missing.
An invalid TC_MEAS or RTD_MEAS unit
parameter (not CEL or FAR) was received, or
an invalid PRES_UNIT or ISO_PRES_UNIT
parameter was received.
An invalid RANGELCK parameter was
received.
RANGELCK ON was received when the
CL3001 is not in Volts mode.
An invalid RTD_TYPE parameter was
received.
An invalid TC_REF parameter was received.
An invalid TSENS_TYPE parameter was
received.