Rtd Test - Omega CL3001 User Manual

Lab calibrator
Table of Contents

Advertisement

1. Disconnect any test leads from external devices.
2. Select voltage output on the primary display as described in
section 4.1
3. Select current input on the isolated display as described in
section 5.2. Select the isolated loop power option.
4. Connect the transmitter as shown in figure 22.
5. Test and calibrate the transmitter per the manufacturer's
instructions.
7.4

RTD Test

1. Disconnect any test leads from external devices.
2. Select RTD input on the primary display as described in section
4.3. Select the RTD type which corresponds to the RTD being
tested.
3. Connect the RTD as shown in figure 23.
4. Test the RTD per the manufacturer's instructions.
Figure 23 - RTD Test Application
35