Siemens SITRANS LR560 Operating Instructions Manual page 57

Radar transmitters
Hide thumbs Also See for SITRANS LR560:
Table of Contents

Advertisement

TVT Shaper
Note: Double click on each axis to see the Xscale and Data Scale values. Right-click or
Left-click on the axis and drag to reposition the scale.
This feature allows you to manually adjust the TVT to avoid false echoes caused by
obstructions. (For an explanation see
Open the menu Device – Echo Profile Utilities and click on the tab TVT Shaper
TVT
Turn Shaper Mode On.
Click on Measure to refresh the echo profile and load the current TVT from the
device.
Change the position of the cursor on the TVT using the Point+ and Point– buttons:
raise and lower the TVT using Offset+ and Offset–.
Alternatively, enter values for Point and Offset directly into the dialog boxes.
Click on Transfer to Device.
Note: If the TVT Shaper tab is closed without transferring new values to the
device, the new values will appear in the PDM list even though they have not been
transferred to the device.
A5E34647946
Shaper Mode
cursor
echo
profile
SITRANS LR560 (mA/HART) – OPERATING INSTRUCTIONS
on page 101.)