Localized information
For the United States and Canada
For the United States and Canada, a CSA or UL-listed power cord must be used.
The connector must be compliant with NEMA 1-15P (NEMA; National Electrical
Manufacturers Association).
● 120 V/240 V supply voltage
A flexible cable with UL approval and CSA marking must be used. In addition, the cable
must exhibit the following properties:
– SPT-2 or SVT with three wires
– At least 18 AWG conductor cross-section (AWG: American Wire Gauge, US standard
– Max. length of 4.5 m
– Connector 15 A, min. 125 V
For countries other than the USA and Canada
● 230 V supply voltage
This device is equipped with a safety-tested power cord. If you choose not to use this
cable, you must use a flexible cable of the following type: At least 18 AWG (0.82 mm
conductor cross-section and 15 A/250 V connector. The cable set must conform to the
safety regulations of the country in which the devices are installed, and bear the
prescribed markings in each case.