M5-CYG2P-M-01_Iss6.doc
2.
Wipe clean the step wedge then
add some fresh couplant to the
step wedge
3.
Place the Probe on the sample
and verify the gauge can get a
thickness value.
4.
Press the Menu key
5.
Press the Up Select key to get to
the Calibrate Two Point item
6.
Press the Menu key to start
calibration
7.
Start with the Thick sample –
Maximum thickness.
While holding the probe firmly
on the thick sample, and while a
steady thickness measurement
is displayed..
Cygnus 2+ Operating Manual
33
Need help?
Do you have a question about the M5-C2P and is the answer not in the manual?
Questions and answers