Cygnus M5-C2P Operating Manual page 31

Multi-mode ultrasonic thickness gauge
Table of Contents

Advertisement

M5-CYG2P-M-01_Iss6.doc
2.
Place the Probe on the sample
and verify the gauge can get a
thickness value.
3.
Press the Menu key
4.
Press the Up Select key to get to
the Calibrate Single Point item
5.
Press the Menu key to start
calibration
6.
While holding the probe firmly
on the sample, and while a
steady thickness measurement
is displayed..
7.
...use the Up and Down Select
keys to adjust the thickness to
the required value.
8.
When done press the Menu key
to save the calibration.
Or press the Power / Cancel to
exit without saving the
calibration.
Cygnus 2+ Operating Manual
31

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M5-C2P and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Cygnus 2M5-c2

Table of Contents