5.
Operation
5.3

Adjusting the clearance angle

21
Enlarged detail:
Index marks for
clearance angle
setting
20
24
27
28
Fig. 12
• The index marks (0°, 5° and 10°) for adjust-
ment of the clearance angle (27) are located
on the right side of the knife holder (25).
• There is also an index mark (28) on the right
side of the knife holder base (21) which
serves as a reference point when adjusting
25
the clearance angle.
• When using the knife holder base (non-
orientable), loosen the Allen screw to
release the clamp by turning Allen key No. 4
(24) counterclockwise.
• When using the knife holder base (with
lateral movement), rotate the lever on the
right side of the knife holder base counter-
clockwise.
• Move the knife holder until the index mark of
the desired clearance angle coincides with
the reference line on the knife holder base.
Example:
Enlarged detail showing a clearance angle
setting of 5°.
The recommended clearance angle
setting for knife holder E is approx.
1° - 3°.
• Firmly hold the knife holder in this posi-
tion and rotate the lever (24) or Allen screw
(depending on the knife holder base used)
clockwise to clamp it.