TiX640, TiX660, TiX1000
Users Manual

Optional Accessories

ac adapter with 14-pin LEMO connector
replacement lenses, macro attachments
protective window for lenses, laser protection filter
breakout box
Bluetooth headset
gigabit Ethernet card
Ethernet cable with 8-pin LEMO connector
device tripod
1-8

Technical Description

Functional Principle

The Product is a thermography system designed for the
long-wave infrared spectral range (LWIR) of 7.5 µm to
14 µm. The lens reproduces the object scene to a
microbolometer array with 1024 x 768 and/or 640 x 480
pixels. The electrical signal of the detector array is
processed further by the internal electronics. In this, the
electronics comprise all functions required for camera
operation, such as actuation of the microbolometer array,
A/D conversion, offset and gain correction, pixel correction,
as well as actuation of the different interfaces.
Operation is optionally performed directly using the
integrated keyboard (joystick, focus rocker switch, function
buttons) or using optional interfaces. RS232, Ethernet (GigE
Vision), or USB2.0 are available as interfaces. Specific
software packages are required in order to use these
interfaces.

Description of the Functional Units

Lens

The camera lens collects the IR radiation in the field of
view emitted by the measuring object and reproduces this
IR radiation on the detector array. Field of view (FOV)
and resolution/measuring spot size (IFOV, Instantaneous
Field of View) are determined at the same distance to the
focal distance f of the used lens.