Technical Description
Functional Principle
The Product is a thermography system designed for the
long-wave infrared spectral range (LWIR) of 7.5 µm to
14 µm. The lens reproduces the object scene to a
microbolometer array with 1024 x 768 and/or 640 x 480
pixels. The electrical signal of the detector array is
processed further by the internal electronics. In this, the
electronics comprise all functions required for camera
operation, such as actuation of the microbolometer array,
A/D conversion, offset and gain correction, pixel correction,
as well as actuation of the different interfaces.
Operation is optionally performed directly using the
integrated keyboard (joystick, focus rocker switch, function
buttons) or using optional interfaces. RS232, Ethernet (GigE
Vision), or USB2.0 are available as interfaces. Specific
software packages are required in order to use these
interfaces.
Description of the Functional Units
Lens
The camera lens collects the IR radiation in the field of
view emitted by the measuring object and reproduces this
IR radiation on the detector array. Field of view (FOV)
and resolution/measuring spot size (IFOV, Instantaneous
Field of View) are determined at the same distance to the
focal distance f of the used lens.