Extras Submenu - Fluke tix640 User Manual

Thermal imager
Hide thumbs Also See for tix640:
Table of Contents

Advertisement

TiX640, TiX660, TiX1000
Users Manual
 Caution
The thermal imager is equipped with a highly
sensitive uncooled microbolometer detector.
The object radiations occurring in
thermographic practice do not induce any
changes to the detector. However, overloads
exceeding a black body radiation of 1,000°C
in the measuring ranges 1 and 2, as well as
2,000°C in measuring range 3 may result in
irrevocable damages and thus must be
avoided.
When using the thermal imager for mobile
applications and in the event of a constantly
changing scene, we recommend deactivating
the automatic measuring range setting (factory
setting).
3-38
Note

Extras Submenu

The settings that can be selected in the "Extras" menu
item activate the following functions:
• Premium
A camera-internal compensation
(NUC) is performed before each
saving.
• Special
Saving is performed with activated
image filter.
• Super
Resolution
To save the image, the hardware-
based microscan procedure is
activated (recording procedure to
increase the geometric resolution).
The thermal image is scanned
optically several times, whereby the
geometric resolution from (1,024 x
768) to (2,048 x 1,536) and/or from
(640 x 480) to (1,280 x 960) IR
pixels is reached by shifting the
hvh224.jpg