System Menu; Units - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

SELECT A MODULE
>SPRT 1
The information for the selected module then appears as follows:
The name of the module is given, followed by its model number. The next line
shows the firmware version number for the module. Following this is a list of
devices contained by the module. Each device is listed with its name, class (i.e.
"INPUT", "OUTPUT", "PRINT", or "COMM"), and the number of channels it
contains. If the list is too long to fit on one screen the UD buttons can be used
to scroll the list. Press EXIT or ENTER to exit.
5.5

System Menu

The SYSTEM menu provides general system functions. The soft-key functions
that appear in this menu are UNITS, TIME, PASSWORD, SYSTEM INFO,
and SYSTEM RESET.
5.5.1

Units

The UNITS soft-key is used to set the units for temperature measurements. The
options are degrees Celsius (C), degrees Fahrenheit (F), or Kelvin (K). A win-
dow appears allowing you to set the units. The units are selected using the
LR buttons and pressing ENTER. Press EXIT to exit.
TCS
2
PRTS 3
COMM 4
SPRT MODULE, MODEL 2560
FIRMWARE VERSION: 1.10
DEVICES:
SPRT INPUT, 2 CHANNELS
5 Soft-Key Functions