Measuring One Channel; Scanning Channels; Displaying Measurement Data - Fluke 1560 Black Stack User Manual

Thermometer readout
Table of Contents

Advertisement

1560 Thermometer Readout
User's Guide
4.5.3
4.5.4
4.5.5
26
EDIT SPRT 01
PROBE SER#: 566-011
CONVERSION: ITS-90
LO RANGE: 4
HI RANGE: 7
R[273]: 25.546738
A[4]: -1.5763669E-4
Press EXIT twice to return to the soft-key menu. The coefficients you entered
can be verified by using the TEST CONV soft-key function (see Section 5.2.3)
to compare calculated temperatures to expected values from a calibration
report.

Measuring One Channel

The 1560 can be set up to measure one channel continuously. To do this the
scan mode must be set to primary channel and the measure mode must be on.
The scan mode is set using the SCAN MODE soft-key function in the INPUT
menu (see Section 5.1.4). Use the LR buttons to set SCAN MODE to
PRIM CHAN and press ENTER. Press EXIT to return to the menu. The mea-
sure mode is set using the MEAS soft-key function in the INPUT menu (see
Section 5.1.1). Use the LR buttons to set MEASURE to ON and press
ENTER. Press EXIT to return to the menu. See Section 4.5.1, Selecting Input
Channels, for information on how to set the primary channel.

Scanning Channels

The 1560 can be set up to measure several channels in sequence. To do this the
scan mode must be set to scan and the measure mode must be on. The scan
mode is set using the SCAN MODE soft-key function in the INPUT menu
(see Section 5.1.4). Use the buttons to set SCAN MODE to SCAN CHAN and
press ENTER. Press EXIT to return to the menu. The measure mode is set us-
ing the MEAS soft-key function in the INPUT menu (see Section 5.1.1). Use
the buttons to set MEASURE to ON and press ENTER. Press EXIT to return
to the menu. The 1560 will scan only channels that are enabled. Channels can
be enabled or disabled using the SCAN CHAN soft-key function in the IN-
PUT menu (see Section 5.1.3).

Displaying Measurement Data

Each measurement appears in the primary measurement window at the top of
the screen and is labeled with the input channel number. The area in the center
of the screen below the primary measurement window can be used to show
measurement data in a variety of formats. To graph measurements, select the
OUTPUT menu, the DISP WINDOW sub-menu, and the GRAPH WINDOW
$