Specifications
Specifications are subject to change without notice.
Specification
Measurement method
Primary compliance method DIN EN ISO 7027
Dimensions (W x D x H)
Weight
Enclosure
Protection class
Pollution degree
Installation category
Power requirements
Operating temperature
Storage temperature
Humidity
Display
Laser
Optical light source
Measurement units
Range
Accuracy
Linearity
Precision
Stray light
Calibration options
Verification options
1
1 mNTU = 0.001 NTU
Details
Nephelometry with the scattered light collected at a 90° angle to the incident light
and 360° around the sample vial.
41 x 28 x 12.5 cm (16 x 11 x 7.7 in.)
2.37 kg (5.23 lb)
IP20
Instrument: III; Power supply: II
2
II
Instrument: 15 VDC, 2 A; Power supply: 100 to 240 VAC, 50/60 Hz
10 to 40 °C (50 to 104 °F)
–30 to 60 °C (–22 to 140 °F)
5 to 95% relative humidity, non-condensing
17.8 mm (7 in.) color touch screen
Class 1 laser product: Contains a non user-serviceable class 1 laser.
850 nm, maximum 0.55 mW
NTU, FNU, TE/F, FTU, EBC, mg/L, mNTU
0 to 1000 FNU, FNU, TE/F, FTU; 0 to 100 mg/L; 0 to 250 EBC
± 2% of reading plus 0.01 NTU from 0 to 40 FNU
± 10% of reading from 40 to 1000 FNU based on Formazin primary standard at
25 °C (77 °F)
Better than 1% for 0 to 40 NTU on Formazin at 25 °C (77 °F)
< 40 NTU: 0.002 NTU or 1% (the larger value); > 40 NTU: 3.5% based on
Formazin primary standard at 25 °C (77 °F)
< 0.01 FNU
®
StablCal
: 1-point calibration (20 FNU) for 0 to 40 FNU measurement range; 2-
point calibration (20 and 600 FNU) for 0 to 1000 FNU (full) measurement range
Formazin: 2-point calibration (20 FNU and dilution water) for 0 to 40 FNU
measurement range; 3-point calibration (20 FNU, 600 FNU and dilution water) for
0 to 1000 FNU (full) measurement range
Degrees: 3-point calibration (20 and 100 mg/L and dilution water) for 0 to 100 mg/L
(full) measurement range
SDVB: 3-point calibration (20 FNU, 600 FNU and dilution water) for 0 to 1000 FNU
(full) measurement range
Custom: 2- to 6-point custom calibration for a measurement range of 0 FNU to the
highest calibration point.
Glass verification rod (secondary turbidity standard) < 0.1 NTU, StablCal or
Formazin (0.1 to 40 NTU)
1
or mFNU
English 3