Hach TU5200 User Manual page 35

Laboratory laser turbidimeter
Hide thumbs Also See for TU5200:
Table of Contents

Advertisement

Message
The sample ID list full. Data
has not been added
Documents/Videos
The document does not
show on the display.
Not enough memory to store
the document.
Cannot delete the
document.
Cannot store the document.
Cannot open the manual.
The video format is not
supported.
The video does not play.
Ethernet/LAN connection
Network authentication
failed
LAN Connection failed
Automatic retrieval of the
local IP address failed!
Setting IP-address failed!
Setting subnet-mask failed!
Setting default gateway
failed!
Mounting network drive
failed!
FTP connection failed!
Verify network configuration
Files did not save to the
network share.
No connection to the web
server.
RFID
Standard expired. Continue? Use standards that are not expired.
Primary standard is write
protected.
Measured value is out of
nominal value range.
No act. value on RFID tag.
Read standard?
Remove the sample IDs that are not used. Add a new sample ID.
Make sure that the document is saved as a PDF.
Delete other documents, then add the document again.
Contact technical support.
Contact technical support.
Make sure that the document format is PDF.
Make sure that the video format is AVI (Codec XVID) with a resolution of 640 x
480 pixels. Make sure that the video size is not more than 10 Mb.
The video file is corrupt.
Enter the correct username and password. Make sure that the network settings
are correctly configured.
There is no connection to the DHCP server. Contact the network administrator.
If the DHCP server is not enabled, manually enter an IP address for the
instrument.
Contact the network administrator. If necessary, contact technical support.
Make sure that the correct permissions are set to the network share directory.
Try to connect the instrument to a LAN again later. Make sure that the instrument
is connected to the network.
The verification standard value cannot be recorded on a calibration standard.
The measured value is not within range for this standard. The 10 NTU standard
range is 5 to 15 NTU. The < 0.1 NTU glass verification rod range is 0 to 0.1 NTU.
Use the correct standard. Make sure that the standard is not expired. Measure the
standard again.
Measure the verification standard first on an instrument that was recently
calibrated.
Solution