Put the vial in the instrument
Personal injury hazard. Never remove covers from the instrument. This is a laser-based instrument and
the user risks injury if exposed to the laser.
Personal injury hazard. Do not look into the vial compartment when the instrument is connected to
power.
Keep the lid closed to keep contamination out of the vial compartment.
1. Log in to the instrument as follows:
• Put an operator RFID tag in front of the RFID module or
• Push Login. Select the applicable operator ID, then push Select.
2. Select the sample ID as follows:
• Put the sample RFID sticker on the sample bottle in front of the RFID module or
• Push Sample ID. Select the applicable sample ID, then push Select.
Note: To add sample IDs to the instrument, refer to
3. Clean the sample vial with a no-lint cloth to remove contamination.
4. Dry the external surfaces of the vial with a no-lint cloth. Make sure to dry the bottom of the vial.
5. Put the sample vial in the vial compartment. Refer to the illustrated steps that follow.
20 English
C A U T I O N
C A U T I O N
N O T I C E
Add sample IDs
on page 14.