SPeCIfICATIOnS
PRODUCT SPeCIfICATIOnS
Weight
shipping Weight
Lifting Capacity
envIROnMenTAl COnDITIOnS
Conditions
Normal Use
storage & Transport
eleCTRICAl SPeCIfICATIOnS
Designed for:
Input service
Current Draw
Maximum Momentary
Current Consumption
Voltage Output to
Actuators
electric shock
Protection
Tabletop Applied Part
Ingress Protection
rating
Made of Operation
Fuse Specifications
280-325 lbs. (127-148 kg.)
372-417 lbs. (169-189 kg.)
550 lbs. (250 kg.)
Temperature
50° (10°C) to 104° (40°C)
-20° (-29°C) to 135° (57°C)
North America
120 VAC/15 amp/60 Hz
5.8 amps
6.0 amps
24 VDC
Class 1 equipment
Type B Applied Part
IPX0
Intermittent Operation
MAX 2 minutes ON
18 minutes off
M 10AL 125V
Humidity
20% to 60% rH
20% to 95% rH
europe
230 VAC/10 amp/50/60 Hz
2.9 amps
3.0 amps
24 VDC
Class 1 equipment
Type B Applied Part
IPX0
Intermittent Operation
MAX 2 minutes ON
18 minutes off
M 10AL 125V
Atmospheric Pressure
98 to 105 kPa
98 to 105 kPa
14