6 Set up with the display and adjustment module PLICSCOM
Service - Adjustment unit
Service - Language
Service - SIL
Service - HART mode
36
Default setting
Like basic adjustment, but in addition, special parameters are reset to
default values.
6)
Peak value indicator
The min. and max. distance values are reset to the actual value.
In this menu item you select the internal arithmetic unit of the sensor.
Unit of measurement
m(d)
The sensor is already set to the ordered national language. In this
menu item you can change the language. The following languages are
available, e.g. in software version 3.50:
Deutsch
English
Français
Espanõl
Pycckuu
Italiano
Netherlands
Japanese
Chinese
Language
German
The functional safety is already activated Ex factory for instruments
with SIL qualification. For instruments Ex factory without SIL qualifica-
tion, the functional safety must be activated by the user for applica-
tions according to SIL via the indicating and adjustment module. The
SIL factory setting cannot be deactivated by the user.
The activation of SIL has the following impact:
In the menu item "Failure mode" under "Current output", the
parameters "Hold value" and "20.5 mA" are blocked
In the menu item "HART mode", the function "Multidrop" is blocked
Note:
For such applications, it is absolutely necessary to take note of
"Safety Manual".
HART offers standard and multidrop mode.
The mode "standard" with the fixed address 0 means outputting the
measured value as a 4 ... 20 mA signal.
Special parameters are parameters which are set customer-specifically on
6)
the service level with the adjustment software PACTware.
VEGAPULS 63 • 4 ... 20 mA/HART - four-wire