13.5.2 Procedure

Follow this procedure:
1. Push the navigation pad to display the menu system.
2. Select
(Color) and push the navigation pad. This displays a submenu.
3. Use the navigation pad to select a different palette.
4. Push the navigation pad to confirm and exit the menu mode.

13.6 Changing the measurement parameters

For accurate measurements, it is important to set the measurement parameters:
• External IR window compensation.
• Object distance.
• Atmospheric temperature.
• Relative humidity.
• Reflected temperature.
• Emissivity.
You can set the measurement parameters globally. You can also change the Emissivity,
Reflected temperature, and Object distance parameters locally for a measurement tool.
For more information, see section 15.5 Changing the measurement parameters, page 71.