GE DigitalFlow XMT868i Startup Manual page 49

Ultrasonic flow transmitter for liquids
Table of Contents

Advertisement

2.4.3c
Pipe OD
5. To select the appropriate Pipe OD Unit type (see Table 16 below), scroll to the right side of the screen, and use
the up and down arrow keys to step through the list. Press
or circumference on the left side and press
Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the transducer
installation site. The data may also be obtained from standard pipe size tables found in Sound Speeds and Pipe Size
Data (GE document 914-004).
inch
feet
in/PI = pipe circumference in inches
ft/PI = pipe circumference in feet
6. Use the arrow keys to enter the known Pipe Wall Thickness (in inches or mm) and press
Note: If you are using clamp-on transducers, skip to step 9 below.
2.4.3d
Path and Axial Lengths
7. To enter the Path Length:
a. Use the [] arrow key to highlight the path length unit type at the right of the screen.Then use the [] and
[] arrow keys to scroll to the desired unit type.
b. Use the [] arrow key to return to the numeric entry on the left, and enter the path length of the ultrasonic
signal. Press
Note: If a spoolpiece was ordered with the meter, the transducer signal path length
axial length
(L)
For on-site transducer installations, refer to Appendix C, Measuring P and L Dimensions, for instructions.
8. In the same manner, enter the appropriate Axial Length L unit type and axial length of the ultrasonic signal, and
press
. Skip to step 10 on page 42.
[Enter]
2.4.3e
Pipe Lining
9. Scroll to the appropriate choice to indicate whether you have a Pipe Lining.
a. If you have selected Yes, then select the Lining Material from the supplied list.
If
is selected, enter the lining sound speed, and press
OTHER
b. Then enter the lining thickness and press
DigitalFlow™ XMT868i Startup Guide
Table 16: Available Pipe OD Units
English
.
[Enter]
are engraved on the flowcell and/or are included in the documentation supplied with the meter.
. Then enter the known pipe outside diameter
[Enter]
.
[Enter]
mm = millimeters
m = meters
mm/PI = pipe circumference in millimeters
m/PI = pipe circumference in meters
[Enter]
.
[Enter]
Chapter 2. Initial Setup
Metric
[Enter]
and the transducer signal
(P)
.
.
41