Linearity
Note:
Caution!
Adjusting External Linearity
Checking the Linearity
The linearity must be checked according to the Standard Operating Procedure
WKD-038.
-
Check the linearity of the balance/scale for the entire weighing range in 4-g
steps.
-
Compare the values displayed with the tolerance ranges given in the „Table of
Calibration/Adjustment Data" on pages 42-43.
-
If the errors exceed the permissible tolerance ranges, the linearity must be
adjusted.
Adjusting the Linearity
You can adjust the linearity on these balances/scales only using the SARTORIUS
MC1 Server Software (Version 4.9 and later) or with the SARTOCAS program
(Version 1.44 and later).
-
Activate the BPI mode (see Page 15).
-
Adjust the linearity with the SARTORIUS MC1 Server (Version 4.9 and later) or
with the SARTOCAS program for PCs and laptops.
(Please see the program description for instructions on this automatic proce-
dure!).
| 39