Funktion Service-Switch
10 |
2
Aut_4106.ipg / CP_bpi_sch.eps
Aut_4106.ipg / CP_bpi_t_sch.eps
Caution!
Left
Right
open
closed
Note:

Access Lock Switch

The access lock switch (2) is located on the rear panel of the balance/scale.
On models CP622, CP4202S, CP3202S, CP3202P, CP2202S, GP5202, GP3202, CP8201,
CP6201, CP4201, CP2201, GP8201, CP423S, CP323S, CP323P, CP153, GC2502,
CP225D, CP324S, CP224S, CP124S, CP64, GC1603, GC803S, GC803P, the opening on
the left is intended for the access lock switch (2) (see fi gure on the left).
With the CP225D it´s in an external electronics box
On models CP34001S, CP34001P, CP34000, CP16001S, CP12001S, the opening on
the left is intended for the access lock switch (2) (see fi gure on the left).
2
Always leave the access lock switch (see left) in its left position. This switch
should only be operated on models verifi ed for use in legal metrology!
Depending on the model (verifi able or standard), the switch has two functions.
1. To unlock the menu to change the menu code settings „ 1 9 2 „ the external
span adjustment on verifiable balances/scales.
External adjustments are generally possible on standard models. It is not neces-
sary to open or close the switch.
2. After the menu is activated „ 8 1 2 „ (read only parameter), it can be released
by moving the switch (2) read only (right position) or for changing the settings
(left position).
On verifi ed models, span adjustment can be carried out with external weights
without the service software.
Procedure:
1. Open access lock switch
2. Activate menu „ 1 9 7 „
3. Press the q key
4. Adjust the balance/scale
5. Close access lock switch!!