4.3.3

Test Lead Connection

The measurement leads are connected using the four binding posts on the
left side of the front panel as shown below.
Connect the two red leads to terminals C1 and P1. Connect the two black
leads to terminals C2 and P2.
Any drop in the voltage on the load terminals is measured between the two
"voltage" (V) leads, P1 and P2. The current leads (C1 and C2) can deliver
current from 10mA to 10A.
4.4

Repetitive Measurements

1. Connect the Kelvin clips to the first object. Press START/STOP.
2. Disconnect the clips. The measurement stops and result is displayed.
3. Connect the clips to the second object. The measurement restarts
automatically. Repeat as necessary.
4. After the last measurement, press START/STOP again.
NOTE: This operation is valid only if all of the objects to be measured have
substantially the same value and all of the measurements are made in the
same range.
Micro-Ohmmeter Model 6240
I
U
17