Oxi 330i/340i
Sensors
AutoRange
Measuring ranges/
resolutions with
CellOx 325 D. O. probe
Measuring ranges/
resolutions with
DurOx 325 D. O. probe
4.2.5 Selecting the measuring range, AutoRange
You can use the DurOx 325 or CellOx 325 sensors. If the automatic
recognition function recognizes the DurOx 325 sensor, the measure-
ment is carried out at a lower resolution. The CellOx 325 sensor mea-
sures with a resolution that is a factor of 10 higher.
Each of the measuring parameters has 2 measuring ranges available.
AutoRange causes the measuring instrument to change to measuring
range II if measuring range I is exceeded, and also vice versa.
To switch the automatic measurement range selection on/off, see sec-
tion 4.6 C
ONFIGURATION
ARng appears on the display.
TDS UpH
O
S
1
7
9 2
Baud
Time
8
1
7
Day.Month No.
Year
Ident
Tref25 Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
Measuring range I
Resolution
Measuring range II
Resolution
Measuring range I
Resolution
Measuring range II
Resolution
. When the AutoRange function is switched on,
Sal
mV/pH
%
mg/l
S/
m
cm
mbar
M
cm
1/ cm
Sal
6
°
C
/
%
K
TP
°
F
Auto Store
ARng
AR
RCL
mg/l
%
0 ... 19.99
0 ... 199.9
0.01
0.1
0 ... 90.0
0 ... 600
0.1
1
mg/l
%
0 ... 19.9
0 ... 199
0.1
1
0 ... 90
0 ... 600
1
1
Operation
mbar
0 ... 199.9
0.1
0 ... 1250
1
mbar
0 ... 199
1
0 ... 1250
1
17