What to do if...
Error message E7
AR flashes
continuously
Measured value too low
Measured value too
Display
Probe symbol flashes
48
E7
E7
E7
Cause
Membrane damaged
– Membrane damaged
– Membrane head not screwed on
tight enough
Cause
No stable measured value
– Membrane contaminated
Cause
– Insufficient
flow
Cause
high
– High amount of dissolved
substances
– Air bubbles bump on the
membrane with high velocity
– The carbon dioxide pressure
is too high (> 1 bar)
Cause
to
to
to
to
– Time-out of the interface
Cause
– Calibration interval expired
Remedy
– Regenerate probe
– Screw membrane head tight
Remedy
– Clean membrane
Remedy
– Provide flow to the probe
Remedy
– Correct solubility function using
the salinity equivalent
– Avoid direct flow to the
membrane
– Measuring not possible
Remedy
– Check the instrument that is
connected
Remedy
– Recalibrate the measuring
system
Oxi 330i/340i