wtw Oxi 330i Operating Manual page 20

Handheld meter
Hide thumbs Also See for Oxi 330i:
Table of Contents

Advertisement

Operation
Printing the calibration
record (340i)
Sensor evaluation
20
The calibration protocol contains the calibration data of the current cal-
ibration. You can transmit the calibration protocol to a printer via the se-
rial interface (see page 33).
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.5.3
before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.99
14:19
Device No.: 12345678
CALIBRATION 02
Cal time: 02.03.99 / 14:19
Cal interval: 14d
OxiCal
Tauto AR
Relative Slope: 0,88
Probe:
+++
After the calibration, the measuring instrument evaluates the current
status of the sensor against the relative slope. The evaluation appears
on the display. The relative slope has no effect on the measuring accu-
racy. Low values indicate that the electrolytic filling will soon be deplet-
ed and the probe has to be regenerated.
Display
E3
Eliminate the error according to
chapter 6 W
HAT TO DO IF
Relative slope
S = 0.8 ... 1.25
S = 0.7 ... 0.8
S = 0.6 ... 0.7
S < 0.6 or S > 1.25
...
Oxi 330i/340i

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Oxi 340i

Table of Contents