pH/ION 7320
ba75935e04
12/2017
8.3
Selecting the measuring method
The following methods are supported:
Standard addition
Standard subtraction
Sample addition
Sample subtraction
Blank value addition
1.
Connect the ISE combination electrode to the meter.
The pH/ISE measuring window is displayed.
2.
If necessary, select the ISE display (unit, mg/l) with <M>.
3.
If necessary, measure the temperature of the test sample with a ther-
mometer.
4.
Open the ISE menu with <MENU/ENTER>.
5.
Thoroughly rinse the electrode with distilled water.
6.
Adjust the temperature of the standard solutions.
7.
Select Method with <> <> and confirm with <MENU/ENTER>.
8.
Select a method with <> <> and confirm with <MENU/ENTER>.
ISE
Calibration
Man. temperature
ISE setup
:
Method
Start method
Back
9.
Select Start method with <> <> and confirm with <MENU/
ENTER>.
Measurement with the selected method begins (see section 8.3.1
S
TANDARD ADDITION
BLANK VALUE CORRECTION
:
25 °C
Standard addition
12.12.2012
12:12
, page 54 ... section 8.3.5 S
(B
LANK VALUE ADDITION
Ion concentration
TANDARD ADDITION WITH
), page 63).
53