4.5.5 Selecting the measuring method

The following methods are supported:
! Standard addition (Std. add.)
! Standard subtraction (Std. sub.)
! Sample addition (Sample add.)
! Sample subtraction (Sample sub.)
! Blank value addition (Blank add.)
1
Perform the preparatory activities according to section 4.5.1.
2
In the measured value display, select the ISE measuring win-
dow with m.
3
Open the ISE menu with g.
4
Thoroughly rinse the electrode with distilled water.
5
Adjust the temperature of the standard solutions.
6
Select Method with 82 and confirm with g.
7
Select a method with 82 and confirm with g.
ISE
Calibration record
Unit
Method
Start method
8
If necessary, change the unit of the measured value display
with 82 and g in the Unit menu.
9
Select Start method with 82 and confirm with g.
Measurement with the selected measurement begins (see
section 4.5.6 ... section 4.5.10).
mg/l
Std. add.
Operation
45