Device Clear And Selected Device Clear Check - Racal Instruments 1994 Service Manual

Universaltimer/counters
Table of Contents

Advertisement

4.4.6
Device Clear and
Selected
Device Clear
Check
4. 4.
6.1
Test as
shown
below:
Action
HP-85 Code
Your
Controller
Set the
1994
to
the
OUTPUT
715;
Total
A
by B
mode
by sending
the
instrument's
listen
"TA"
address, followed
by
device-dependent
command TA
Send
the
DCL
command
true
CLEAR
7
a.
Verify that the function indicated
on
the
front
panel
changes
to
FREQ A
b.
Test
as
shown
below:
Action
HP-85 Code
Your
Controller
Reset the 1994
to
the
OUTPUT
715;
Total
A
by
B mode
by
sending the
instru-
"TA"
ment's
listen
address,
followed by device-
dependent
command
TA
Send
the
SDC
message
true
CLEAR
715
c.
Verify that the function indicated on the
front panel
changes
to
FREQ A
4-16