Remote And Local Check - Racal Instruments 1994 Service Manual

Universaltimer/counters
Table of Contents

Advertisement

4.4.2
Remote
and Local
Check
4.4.
2.1
Now
perform
the following
procedure:
a.
Power-on
the counter.
Verify that the
REM, ADDR,
and
SRQ
LEDs
flash
on
and
off
once.
If
the
indicators
do
not
flash
or
flash
continuously, there
is
a
fault
on
the
GPIB PCB.
b.
Test
as
shown
below:
Action
HP-85 Code
Your
Controller
Send
the
REN
command
true,
together with
the 1994's
listen
address
REMOTE
715
c.
Verify that the
REM
LED
lights
d.
Test
as
shown
below:
Action
HP-85 Code
Your
Controller
Send
the device-
dependent
command
CK
OUTPUT
715;
"CK
n
e.
Verify that the
ADDR
LED
lights
and
that the
Check mode
is
selected
f.
Test
as
shown
below:
Action
HP-85 Code
Your
Controller
Send the
1 994's listen
address followed
by
the
GTL
command
LOCAL
715
g.
Verify that the
REM
LED
is
off.
The
ADDR
LED
will
also extinguish
if
the controller
automatically
transmits the
UNL
(unlisten)
command
true
as with the
HP-85
4
-
15
!

Advertisement

Table of Contents
loading

Table of Contents