Quick start guide for
Altrix Precision Temperature Management System
Disinfecting the internal water circuit and hoses (cont.)
8. Press
and hold the Stand-by button.
9. Tap
the Manual mode icon.
10. Tap Confirm.
11. Set the water target temperature to 25.0° C (77.0° F).
12. Tap Confirm.
13. Run the controller for 20 minutes.
14. After 20 minutes, turn the controller off by pressing and holding the
Stand-by button for two seconds.
15. Unplug the power cord from the wall outlet.
16. Place the controller over a floor drain.
17. Remove the reservoir. Pull forward at an angle, and lift out the
reservoir.
18. Remove the bottom hose end from the hydraulic connector adapter in
the reservoir lid by pushing down on the collar.
19. Empty water from the reservoir, dispose of the water
per hospital protocol.
8001-069-002 Rev B.2
Refer to the Altrix Precision Temperature Management System Operations Manual for additional information.
20. Pull up on the controller drain plug (Figure 31)
to open the drain.
21. Make sure that all 3 hoses remain raised above the connection ports
for draining.
22. Allow the controller and hoses to drain for a minimum of two minutes.
23. Push down on the controller drain plug to close the drain.
24. When the controller and hoses are drained, continue to Rinsing the
internal water circuit and hoses.
Figure 31