Exercise 4 – Optimize Acquisition parameters
For this exercise you optimize a mixture of four sulfonamide compounds.

Task 1. Use Optimizer to optimize acquisition parameters

Optimizer helps you optimize acquisition parameters. Specifically, it
automates the selection of the best precursor ions, the optimization of the
fragmentor voltage for each precursor ion, selection of the best product ions,
and optimization of collision energy values for each transition for a list of
compounds you specify.
To do this task, you first need to create the method
5. Find optimum collision energy for MRM acquisition"
acquire the data file.
Steps
1 Start the MassHunter Optimizer
program.
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
Task 1. Use Optimizer to optimize acquisition parameters
Detailed Instructions
• Double-click the Optimizer icon.
Exercise 4 – Optimize Acquisition parameters
iiiSulfamix MRM_
on page 30
.
You do not need to
Comments
. • If you are optimizing peptides, use
the Optimizer for Peptides program.
10
.m
in
"Task
59