Agilent Technologies 6460 Manual page 25

Masshunter workstation - data acquisition
Hide thumbs Also See for 6460:
Table of Contents

Advertisement

Steps
3 Set up a qualitative method to
integrate and extract product ion
spectra.
Use the data files
SulfamixPI_xx.d, where xx is
the collision energy, or your own
data files, iiiSulfamixPI_xx.d.
Open Method Explorer and
Method Editor.
Use TICs set up for MS/MS,
product ion and each of the
precursor ions 271, 279, 285,
311.
Make sure the MS/MS
integrator has been selected and
the maximum number of peaks
has been limited to the largest
100 peaks.
Add the ability to integrate and
extract peak spectra to the file
actions run upon data opening.
Save the changes to the current
method.
Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
Detailed Instructions
a Click the Open Data File icon in the
toolbar.
b Select SulfamixPI_15.d.
c Clear the Run File Open Actions from
Specified Method check box, and
click Open.
d Make sure the Method Explorer and
the Method Editor windows are
displayed; otherwise, click the Method
Explorer and then Method Editor
icons.
e In the Chromatogram section in the
Method Explorer window, select
Define Chromatograms.
f Delete any existing chromatograms in
the Defined Chromatograms list.
g Select TIC from the Type list in the
Define chromatograms section.
h For MS level, select MS/MS.
i Mark the Do cycle sum check box.
j For Scans, select Product ion.
k For Precursor ion m/z, type
l Click the Add button.
m Repeat steps j and k for each ion.
Exercise 1 – Develop an acquisition method
Task 4. Determine product ion masses
Comments
• The Qualitative Analysis program
should already be open and contain
iiiexercise 1.m as the method.