Seconds Per Division Field (Timing only)
Seconds Per Division Field (Timing only)
When the analyzer Type field
in
the Configuration menu is set to Timing, the
analyzer uses its own internal clock. In this mode, the X-axis of the
waveform display is measured
in
seconds per division (sec/Div).
You use the sec/Div field to select the seconds per division resolution of the X
axis. The range of the sec/Div field is
1
ns/Div to 1.0 ks. You set the sec/Div
field by either touching the sec/Div field and rotating the knob, or by
touching the sec/Div field twice and then using the pop-up keypad. When
using the knob to set the sec/Div, the value
will
change
in
a 1-2-5 sequence.
By adjusting the sec/Div, you can zoom
in
to view a desired part of the
display.