The Format Menu
Label Polarity Fields
Label Polarity Fields
The Label Polarity fields are used to assign a polarity to each label. The
default polarity for all labels is positive ( + ) . You change the label polarity by
touching the polarity field, which toggles the polarity between positive
(+)and negative ( -).
When the polarity is inverted, all data as well as bit pattern specific
configurations used for identifying, triggering, or storing data reflect the
change of polarity.
In a timing analyzer with the data inverted, the waveform display does not
change.
Polarity field
(
Format 1
State A quisi lion Mode
Full Chann l /4K
Memory/lOOMHz,
Lab3
Lab4
Lab5
Lab6
Lab?
Lab8
Polarity Field
4-26
:: II:
________ uuuu
15 ' ' ' 87 ' ' ' ' 0
· · · ********
I / · ·
· .. ·
~~~~~~~~
I
" " " " " '
-
..
- .